Functioning
Power board temperature probe open circuited or
disconnected
When this alarm is activated, the "NTC" alarm appears
on the display.
This alarm indicates that the power board probe is faulty
or disconnected.
Call the technical support to replace the board
Safety thermostat
When this alarm is activated, the "TS" alarm appears
on the display.
This indicates that the temperature in the chamber
exceeds 512°C; the oven heating system is turned off.
Reset the safety thermostat as described in the relevant
paragraph.
Chamber overheating
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the temperature in the chamber
exceeds 480°C; the oven heating system is turned off.
Turn off the oven and wait for the chamber to cool down.
Try again and call the technical support if the alarm
occurs again.
Chamber temperature probe open circuited or dis-
connected
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the chamber temperature probe is
open circuited or disconnected; the oven is turned off.
Call the technical support to check and/or replace the
probe.
Chamber temperature probe short circuit
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the chamber temperature probe is
short circuited; the oven is turned off.
Call the technical support to replace the probe
Technical compartment temperature probe open
circuited or disconnected (only for American market)
When this alarm is activated, the "TC2" alarm appears
on the display
This indicates that the technical compartment tempera-
ture probe is open circuited or disconnected; the oven
is turned off.
Call the technical support to check and/or replace the
probe.
EN - 39