must be replaced. Check again with the new connecting tubing (repeat steps
2a-2c).
During connection and disconnection of the connecting tube, do not bend the
luer fitting of the device, to prevent damage. Furthermore, make sure that the
connecting tubing is held on the connector (not at the tubing).
If the vacuum valve is used, do the following test prior to use:
3a) Inflate the low pressure cuff of a tracheal tube with the Cuff Pressure Gauge to
40 cmH
O (picture 4).
2
3b) Attach the inflation line to the Luer nozzle of the red vacuum valve (picture 5, 6).
3c) Squeeze the inflation bulb several times; the low pressure cuff has to deflate.
3d) Squeeze the inflation bulb until it does not return to its original position; the low
pressure cuff must be completely empty now (picture 6).
3e) Disconnect the inflation line from the vacuum valve. Connect the inflation line
to the Luer nozzle of the Cuff Pressure Gauge. The needle of the Cuff Pres-
sure Gauge has to show a negative pressure by moving counter-clockwise (pic-
ture 7).
INSTRuCTION FOR uSE
Before use check the low pressure cuff for leaks. Prior to intubation or extubation
withdraw all air from the cuff via a syringe or the vacuum valve. Connect the Cuff
Pressure Gauge (eventually by using the connecting tubing) to the inflation line of
the tracheal tube. Inflate the cuff to a pressure of 60 - 90 cmH
the cuff is in close contact with the tracheal wall. Release the pressure immediately
by pressing the red release valve until the pointer reaches the green area. We
recommend the intra cuff pressure to be above 22 cmH
tion or pneumonia and below 32 cmH
mucous membrane (see publication of J. Stauffer in Respiratory Care, July 1999).
Check and monitor the cuff pressure continuously. Adjust pressure increase via the
red release valve and pressure decrease via the inflation bulb.
wIPE DISINFECTION
Wipe disinfection using conventional surface disinfectants based on alcohol (e.g.,
Incides
N from Ecolab Deutschland GmbH) is suitable for disinfection. When
®
choosing a product for disinfection, a disinfectant with appropriate ranges of action
must be used: bactericidal, yeasticidal, tuberculocidal and virucidal. After the wipe
disinfection, the devices must be inspected for visible contamination. If neces-
sary, repeat the wipe disinfection. After the wipe disinfection, the devices must be
checked in accordance with the section „Functional check".
CAuTION
• Do not immerse the product in liquids.
• The product cannot be automatic or manual cleaned respectively sterilised.
• Do not clean the glass (plastic material) with aggressive agents.

METROlOGICAl INSPECTION

The accuracy ±2 cmH
24 months. If required measuring equipment is not available, the metrological con-
O due to the risk of ischemia of the tracheal
2
O of the Cuff Pressure Gauges must be checked every
2
2
-11-
O. This ensures that
2
O due to the risk of aspira-