Chemical Method
Dichromate / H
SO
2
4
Appendix
Interferences
Persistant Interferences
In exceptional cases, contents, for which the oxidation capacity of the reagent is not
sufficient, can lead to lower results.
Removeable Interferences
Suspended solids in the vial can lead to incorrect measurements and so to avoid
this, it is important to place the vials carefully in the sample chamber as the method
necessitates a build-up of precipitate at the bottom of the vial.
The outer walls of the vial must be clean and dry before the analysis is carried out.
Fingerprints or water droplets on the vial lead to incorrect measurements.
In the standard version, chloride interferes from a concentration of 10000 mg/l. In
the mercury-free version, the disturbance depends on the chloride concentration and
the COD. Concentrations from 100 mg/l chloride can lead to significant disturbances
here. To remove high chloride concentrations in COD samples, see method M130
COD LR TT.
Method Validation
Limit of Detection
Limit of Quantification
End of Measuring Range
Sensitivity
Confidence Intervall
Standard Deviation
Variation Coefficient
Conformity
ISO 15705:2002
According to
ISO 15705:2002
Reactor is necessary for COD (150 °C), TOC (120 °C) and total -chromium, - phosphate, -nitrogen, (100 °C)
b)
EN   Method Reference Book   1.0
COD HR TT / M132
112.81 mg/l
338.43 mg/l
15 g/l
21,164 mg/l / Abs
70.48 mg/l
27.84 mg/l
0.37 %
EN
19