Chlorine HR T / M103
Chemical Method
DPD
Appendix
Interferences
EN
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
The use of reagent tablets in samples with high Calcium content* and/or
high conductivity* can lead to turbidity of the sample and therefore incorrect
measurements. In this case, the alternative reagent tablet DPD No. 1 High Calcium
and reagent tablet DPD No. 3 High Calcium should be used.
*it is not possible to give exact values, because the development of turbidity depends
on the composition and nature of the sample.
Conformity