INSTRUCTION MANUAL CODE 80259 REV 0 January/2011
Fig. 2
Temperature probe connector location.
If you don't use the external probe, the
«bridge connector» must be placed.
Fig. 3
External probe operation.
J
.P. SELECTA, s.a.
Operating without external probe:
The temperature controller shows and controls the mantle temperature. This
temperature is slightly higher than the sample temperature.
This operating mode is appropriate when there's no need of a very accurate
control of the sample temperature.
1. Place the «bridge connector» (see Fig. 2). It normally comes placed by
default.
2. Select the working temperature.
Operating with external probe:
The temperature controller shows and controls the probe temperature im-
mersed in the sample. This allows an accurate control of the temperature at
which the sample is being treated.
- The sample liquid must be compatible with the probe
stainless steel.
- Do not submerge the probe in hard acids.
This operating mode is appropriate when you need a very accurate control of
the sample temperature.
1. Extract the «bridge connector» (see Fig. 2) and connect the external probe
(see Fig. 3).
2. Fix the probe with the help of a bar and a clamp. Then submerge it about
2 cm in the sample.
3. Select the working temperature.
Be sure that the probe is immersed in the fl ask liquid
before switching the mantle on.
If not, the sample temperature will rise in an uncon-
trolled way.

Maintenance

CLEANING:
It is recommended the use of the following products for the cleaning of the
equipment different parts:
Cleaning of casing, plastic and painted metals: Alcohol with cotton or a duster.

Guarantee

This product is guaranteed for one year. The guarantee does not cover damage
caused by incorrect use of causes beyond the control of J.P. SELECTA, S.A.
Any manipulation of the equipment by unauthorized personnel by J.P. SELECTA,
S.A. cancels the guarantee automatically.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - http://www.jpselecta.es
(It can be modifi ed without notice)
ATTENTION!
D A N G E R
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
Page: 8
Tabla de contenido
loading

Este manual también es adecuado para:

Fibroman ht-w 250mlFibroman ht-w 500mlFibroman ht-w 1000mlFibroman ht-w 2000ml30314703031471 ... Mostrar todo

Tabla de contenido