Troubleshooting - Graham Field JOHN BUNN Neb-u-Tyke Bella the Butterfly Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TRoUbLeShooTIng

If your Neb-u-Tyke Bella the Butterfly Nebulizer
Compressor fails to function, consult the following
Troubleshooting guide. If the problem persists, consult
your Graham-Field authorized distributor.
Problem
Device doesn't
operate
Nebulization is
weak
Bella the Butterfly Nebulizer Compressor
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 20 minutes,
auto-shut-down may have been activated by the built-in thermal
protector; ensure that device is OFF for at least 40 minutes before
next usage.
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.