Operation
Use the iC2 only with the corresponding app.
Adjust the brightness as low as necessary during the application and ensure that the light is switched
off after each application.
The iC2 is controlled by operating elements located on the iC2 and in the iC2 app.
The device provides the following functions:
- Create a patient file / Recording of patient data
- Picture recording
- Filter selection
- Storage of pictures
- Three-level brightness setting via the setting menu of the app.
Protecting patient data is users' sole responsibility (e. g. Protection of the smartphone with a pin, data
transmission via a secure connection). Ensure that your smartphone is adequately protected against
access by third parties.
An electrical conducting connection between smartphone, PC and a mains power source is not
permissible during the recording of the picture.
The pictures are stored in a file in the picture database of the smartphone. This file should only be
viewed and changed with the iC2 app.
Carry out periodically a back-up of your data to avoid a loss. A back-up of the data is for example
possible with the iCloud.
Switch off the device
Switch off the device after each application. This is possible through the following procedure:
- Closing the iC2 app causes the device to switch off automatically.
- If there is no Bluetooth
release button isn't pushed, the device will switch off automatically.
Reset the device
If the device does not react, it can be reset by pushing the release button for 30 seconds. This procedure has
no influence on the data of the iC2 app.
Application duration
The device can be used continuously in an ambient temperature of between 10 °C and 30 °C. In an ambient
temperature of between 30 °C and 35 °C, the iC2 can be used for a brief period of < 15 minutes and with a
break of 10 minutes before being used again.
Charging
Charging via USB
Please connect the USB cord, which is attached to the power supply, to the micro USB charging port (3).
Wireless charging
When using the iC2 charging station, insert the device into the charging station until it lies full-facedly.
Pay attention to the fact that no foreign objects are between the device and the charging station.
When replacing the iC2 in the charger, ensure that the charger and the iC2 aren't contaminated.
The setting up and operation of the iC2 charging station is described in a separate instruction of use.
Due to local regulations, the iC2 charging station may not be available in your country.
Charge Indicator
When the iC2 is charged via USB, the LED will flash green.
When the battery is fully charged, the LED will be constant green.
When using the iC2 charging station, the charging procedure is indicated by the charging station.
8/52
connection between the iC2 and the app for five minutes and the double-stage
®
MED 113539
2020-06-02