Replacement of the nebulizer
The nebulizer and other accessories should be replaced according to its instructions.
Replacement of the filter
The filter is to be replaced when it appears discolored or wet.
Troubleshooting
The device does not turn on
• Make sure the plug is firmly fitted to the electrical outlet.
In the event of malfunction (insufficient pressure and flow) turn the unit off for 10 seconds
and turn back on.
After each cleaning, check that the unit is working properly by verifying that air still comes
out of the nozzle when the unit is turned to the "ON" position.
Unauthorized repair voids warranty.
If the device does not work properly after attempting these troubleshooting steps, contact
your authorized product distributor or Philips Respironics Customer Service at (inside the US)
1 800 345 6443, (outside the US) +1 724 387 4000.
Maintenance and repairs
Never open the device. There are no serviceable parts within the unit. The compressor requires
no lubrication or maintenance.
Technical specifications
Electricity supply
120V/60Hz 1.7 Amp
Thermal overload protector
Thermal fuse; functioning temperature 150 C°
Compressor max. pressure
303 kPa (44 psi)
Average flow rate
7 LPM @ 10 psig
Max flow rate
9.4 LPM
Weight
3.3 lbs
Size
4.2 (H) x 6.5 (W) x 6.5 (L) inches
Noise level
58 ±3dBA
Class II device, Type BF device (device with specific protection against electrical hazards).

DOUBLE INSULATED

Reference to standards
Electric safety standards IEC 60601-1
Electromagnetic compatibility according to IEC 60601-1-2
Ambient conditions
Storage conditions
Temperature
MIN -25 °C – MAX 70 °C (-13 °F – 158 °F)
Humidity
MIN 10% RH – MAX 95% RH
Operating conditions
Temperature
MIN 5 °C – MAX 40 °C (45 °F –104 °F)
Humidity
MIN 10% RH – MAX 95% RH
Altitude
From 0 – 2000 m above sea level
Operating atmospheric pressure
Performance may vary depending on altitude above sea level,
barometric pressure and temperature
Symbol glossary
l
ON (power)
~
O
OFF (power)
SN
Serial number
Atmospheric pressure
Manufacturer
IP21
Housing Ingress Protection rate: IP21 (protected from the penetration of solid bodies with dimensions greater than 12 mm.
Protected from the penetration of vertically falling water drops).
Warranty
Respironics, Inc warrants the compressor to be free from defects in materials and workmanship under normal
use and operation for a period of 5 years from date of purchase from Respironics, Inc. The warranty is limited
to repair or replacement, at Respironics, Inc's sole option, of any such component or equipment claimed to be
defective when claim is shown to be bona fide by evaluation by Respironics, Inc. This warranty does not extend
to any components or equipment subjected to misuse, improper operation, accidental damage, or unauthorized
repairs, and is not extended to charges of, or for, labor repairs. All items returned must be properly packaged
and shipped, prepaid, by the product distributor servicing the unit. Respironics, Inc shall not be liable to
purchaser or others for loss of use of equipment or for indirect, or incidental or consequential damages that
might arise.
ENGLISH
ATTENTION
Class II, double insulated
Alternating current
Type BF applied parts
Follow instructions for use
Indoor use only
Temperature limitation
Humidity