Riester ri-thermo fastPRObe Instrucciones De Uso página 51

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 80
a.)
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and landmobile radios, amateur radio, AM and FM radio bro-
adcast and TV broadcast cannot be predicted theoreticallywith accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic
sitesurvey should be considered. If the measured field strength in the location in
which the ri-thermo
compliance level above, the ri-thermo
served to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the ri-thermo
fastPRObe Thermometer.
b.) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
portable and mobile RF communications equipment and the ri-thermo
The ri-thermo
®
magnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ri-thermo
prevent electromagnetic interference by maintaining a minimum distance bet-
ween portable and mobile RF communications equipment (transmitters) and
the ri-thermo
fastPRObe Thermometer as recommended below, according
®
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recom-
mended separation distance d inmetres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is themaximum output pow-
er rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propaga-
tion is affected by absorption andreflection from structures, objects and people.
fastPRObe Thermometer is used exceeds the applicable RF
®
Recommended separation distances between
fastPRObe Thermometer.
fastPRObe Thermometer is intended for use in an electro-
Separation distance according to frequency of trans-
mitter
(m)
150 kHz to 80
MHz
d =12√P
0,12
0,38
1,2
3,8
12
fastPRObe Thermometer should be ob-
®
fastPRObe Thermometer can help
®
80 MHz to 800
MHz
d =1,2√P
0,12
0,38
1,2
3,8
12
®
®
800 MHz to
2,7 GHz
d =2,3√P
0,23
0,73
2,3
7,3
23
51
Tabla de contenido
loading

Tabla de contenido