Guidance and manufacturer's declaration - electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC61000-4-3
80 MHz to 2.5 GHz
ENGLISH
Compliance
level
3 Vrms
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the device, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
3.5
√ P
d=
V
1
3.5
√ P
d=
80 MHz to 800 MHz
E
1
7
√ P
d=
800 MHz to 2.5 GHz
E
1
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site survey,
be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
88
a
should
b