Environment Requirements; Principle And Caution; Principle Of Measurement - Gima OXY-2 Guía De Uso

Ocultar thumbs Ver también para OXY-2:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
23
The product is not suitable for use in continu-
ous supervision for patients.
The problem of overrating would emerge
when the patient is suffering from toxicosis
which caused by carbon monoxide, the de-
vice is not recommended to be used under
this circumstance.

2.3 Environment Requirements

Storage Environment
a) Temperature: -40°C~+60°C
b) Relative humidity: ≤95%
c) Atmospheric pressure: 500hPa~1060hPa
Operating Environment
a) Temperature: 10°C~40°C
b) Relative Humidity: ≤75%
c) Atmospheric pressure: 700hPa~1060hPa

3 PRINCIPLE AND CAUTION

3.1 Principle of Measurement

Principle of the Oximeter is as follows: An experience
formula of data process is established taking use of
Lambert Beer Law according to Spectrum Absorption
Characteristics of Reductive Hemoglobin (Hb) and
Oxyhemoglobin (HbO
Operation principle of the instrument is: Photoelectric
Oxyhemoglobin Inspection Technology is adopted in
accordance with Capacity Pulse Scanning & Recording
Technology, so that two beams of different wavelength
of lights can be focused onto human nail tip through per-
spective clamp finger-type sensor.
Then measured signal can be obtained by a photosensi-
tive element, information acquired through which will be
shown on screen through treatment in electronic circuits
and microprocessor.
ENGLISH
) in glow & near-infrared zones.
2
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Cms50dl35072

Tabla de contenido