Gima CORSIA Manual De Istrucciones página 17

Ocultar thumbs Ver también para CORSIA:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

The CORSIA PROFESSIONAL NEBULIZER is intended for use in the electromagnetic environment specified below. The
customers or the user of the CORSIA PROFESSIONAL NEBULIZER should assure that it's used in such an environment
Immunity Test
Conducted Immunity
EN 61000-4-6
Radiated Immunity
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial
mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and
accurately foreseen.
To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic study of the site should
be considered. If the field intensity measured in the place where the device will be used surpasses the above mentioned
applicable level of conformity, the normal functioning of the device should be monitored. If abnormal performance arises,
additional measures such as changing the device's direction or positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.
The CORSIA PROFESSIONAL NEBULIZER is intended to operate in an electro-magnetic environment where RF irradiated interferences
Recommended separation distance between portable and mobile radio-communication devices and the monitor
are under control. The client or operator of the CORSIA PROFESSIONAL device can help prevent electro-magnetic interference by
keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and the CORSIA
PROFESSIONAL device, as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal output
power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people.
Guidance and manufacturer's declaration – Immunity Emissions
Level indicated by the
EN 60601-1-2
3Vrms 150kHz to 80Mhz
(for non life-supporting
devices)
3V/m 80MHz to 2.5GHz
(for non life-supporting
devices)
Separation distance from the frequency transmitter (m)
150 kHz to 80 MHz
d = [3.5 / V
]
P
1
0,12
0,38
1,2
3,8
12
Compliance level
Electromagnetic environments -
V
= 3 V rms
1
The portable and mobile RF communication
closer to the CORSIA PROFESSIONAL device,
devices, including cables, must not be used
E
= 3 V / m
than the separation distance calculated by the
1
equation applicable to the transmitter
frequency. Recommended separation distance
d = [3.5 / V
d = [12 / E
d = [23 / E
Where P is the maximum nominal output
voltage of the transmitter in Watt (W)
depending on the manufacturer of the
transmitter and the recommended
separation distance in metres (m). The
intensity of the field from the fixed RF
transmitters, as determined by an electro-
magnetic study of the site
lower than the level of conformity of each
frequency interval
It is possible to check for interference in
proximity to devices identified by the
following symbol:
80 MHz to 800 MHz
d = [12/E
]
P
1
0,12
0,38
1,2
3,8
12
17
.
guidance
P
]
1
]
P
from 80 MHz to 800MHz
1
]
P
from 800 MHz to 2.5 GHz
1
, could be
a)
.
b)
800 MHz to 2.5 GHz
d = [23/E
]
P
1
0,23
0,73
2,3
7,3
23
GB
Tabla de contenido