BATTERY RECHARGEMENT
How to charge the battery
The stimulator is supplied together with a rechargeable battery. The batteries are
Nickel Metal hydrate type with peak effi-
ciency without memory effect. When the
battery indicator lights up on the display it
is advisable to recharge the batteries.
Shut off the stimulator and the electrodes;
connect it to the supplied charger (Globus
MA 075D015CE), inserting the plug in the
special outlet (see the side figure). The LED
on the charger is lit when the connection is correctly executed. The first recharge
must be absolutely complete and it has not to be stopped for any reason. A com-
plete recharge takes about 10 hours; should the stimulator accidentally stay con-
nected for longer than (72 hours) the battery will not be damaged.
Only use the charger supplied with the stimulator. Battery should only be replaced
by an authorized centre.
Attention
Do not use a charger different from that supplied with the stimulator.
During the charger the stimulator MUST NOT be connected to the electrodes.
If the charger is connected with the stimulator switched on, it will automatically
switch off.
Should the cable or connector show signs of wear and tear you should replace it
immediately.
15
ELITE MULTI Rev. 2.1