Solvent Data - Heidolph Hei-VAP Value Instrucciones De Funcionamiento Originales

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

12.3 Solvent data

The graph shows the relationship between the pressure and boiling temperature of a selection of solvents.
The temperature difference between the vapor temperature and the cooling medium
should be at 20 K to result in sufficient condensation.
The temperature difference between the heating bath and vapor temperature should be at
20 K to result in a sufficient Distillation rate (dT).
i.e.: Set a vacuum for a boiling point at 40 °C, set the heating bath temperature at 60 °C.
Figure 12-1: Graph
Torr to mbar conversion: [mmHg] ≈ 3/4 [mbar]
Solvent data
Solvents
Acetone
Acetonitrile
Benzene
n-butanol (butyl alcohol)
tert.-butanol (tert-butyl alcohol)
2-Butanone (methyl ethyl ketone)
Chlorobenzene
Cyclohexane
1.2 Dichloroethane
Hei-VAP Value, Hei-VAP Advantage
Total
MW
Boiling
formula
point
[g/mol]
[°C]
C
H
O
58,08
56,5
3
6
C
H
N
41,05
81,8
2
3
C
H
78,11
80,1
6
6
C
H
O
74,12
117,5
4
10
C
H
O
74,12
82,9
4
10
C
H
O
72,11
79,6
4
8
C
H
CI
112,60
132,2
6
5
C
H
84,16
80,7
6
12
C
H
CI
98,96
82,4
2
4
2
Appendix
ΔHvap
Vacuum for a
boiling point at
[J/g]
40 °C
[mbar]
[mm(Hg)]
550
556
387
833
230
173
549
236
177
619
25
19
588
130
98
473
243
182
375
36
27
389
235
176
336
210
158
EN
71
Tabla de contenido
loading

Este manual también es adecuado para:

Hei-vap advantage

Tabla de contenido