Riester ri-thermo tymPRO Instrucciones De Uso página 19

Termómetro timpánico
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 64
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment
Immunity test
Electrostatic
discharge(ESD) IEC
61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency(50, 60
Hz) magnetic field IEC
61000-4-8 U
NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer's declaration-electromagnetic emissions
Compliance
Group 1
Class B
Not applicable
Not applicable
Manufacturer's declaration-electromagnetic immunity
IEC 60601 test level
Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,±15
kV
± 2kV for power supply
lines
± 1kV for input/output lines
± 0.5kV, ±1kV line(s) to
line(s)
± 0.5kV, ±1kV,±2kV line(s)
to earth
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
Electromagnetic environment-guidance
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Compliance level
Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,±15
kV
Not applicable
Not applicable
Not applicable
Not applicable
Voltage dips:
Not applicable
Not applicable
Not applicable
Voltage interruptions:
Not applicable
30 A/m
50 Hz and 60 Hz
EN-17
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%
Mains power quality should be that of a typical
home and professional environment.
Mains power quality should be that of a typical
home and professional environment.
Mains power quality should be that of a typical
home and professional environment. If the user
of the device requires continued operation during
power mains interruptions, it is recommended that
the device be powered from an uninterruptible
power supply or a battery.
The device power frequency magnetic fields
should be at levels characteristic of a typical
location in a typical home and professional
healthcare environment.
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Ri-thermo tympro+

Tabla de contenido