EXTERNAL CLEANING
Clean with a damp cloth using non-abrasive and non-corrosive detergents (neutral pH value).
MAINTENANCE OF THE WATER BLOCK
Depending on the city water quality in your area and the frequency of use of the equipment, the valve inlet
filter should be regularly cleaned as explained in the instruction leaflet attached to the end of this manual.
CAUTION
Regular control of the Water Block and one-way valve
To ensure compliance with subsection 4.6 of EN1717, have the Water Block and the one-way
valve checked at least once a year; in case a sterilizer is attached to the system, this could be
done during the scheduled maintenance visits for the sterilizer.
It is further recommended to check all related devices fitted to the sterilizer to ensure that the
hydraulic connection between demineralizer and sterilizer corresponds to category 2, in
accordance with EN 1717.
WARNING
The Water Block serves as an anti-flooding device. Blocking only takes place if there is
considerable leakage (such as a cut in the tube between the Water Block and the demineralizer).
Blocking does not take place if there is a small leak in the product's hydraulic system.
Osmo
MAINTENANCE
®
AND ONE-WAY VALVE
17