Trouble Shooting; Error Messages - Adam Equipment CORE Serie Manual De Usuario

Idiomas disponibles
  • ES

Idiomas disponibles

EN 
6.0  TROUBLE
6.1  ERROR
MESSAGES 
 
If an error message is shown, repeat the step that caused the message. If the error 
message is still shown then contact your dealer for support. 
ERROR 
DESCRIPTION 
CODE 
ADO
A/D Value too high. 
ADL
A/D Value too low. 
-----
A/D count is not correct 
when turning the 
balance on. 
Unit-E
Unit error. 
Low battery indicator.  Batteries may be flat. 
FAIL
Calibration failure. 
Unstable – machine 
cannot get a stable 
reading. 
 
No Power when turning 
on. 
 
 
 
 
© Adam Equipment 2016
SHOOTING 
 
POSSIBLE CAUSES
 
Weight on the pan when turning 
on.  
Improper calibration of the 
balance. 
Damaged load cell. 
Damaged Electronics. 
Pan is not installed when turning 
on. 
Calibration not correct. 
Damaged load cell. 
Load cell damaged.
Electronics damaged. 
  
Incorrect mass used to calibrate. 
The user calibration is not within 
5% of factory calibration.  
Possible damaged to loadcell. 
Possible damage to the Loadcell. Make sure the balance is on a 
Battery may be flat. DC adapter 
may not be working.  
14
Solutions 
Remove any weight from the 
stainless steel pan.  
Recalibrate. 
Install pan and power back 
on.  
 
Recalibrate. 
Remove any weight from the 
stainless steel pan and try 
rebooting. Try calibrating the 
balance.  
 
Charge the battery. 
Try to recalibrate, check that 
the weight used to calibrate 
matches that of the value of 
the balance.  
flat surface and away from 
vibration.  
Change the batteries. Or use 
the DC adapter 
308669539revB6-Mar17