Vent Valve Installation - Varian Turbo-V300 ICE Manual De Instrucciones

Tabla de contenido
The metallic model is assembled as shown in the
figure.
Two 6 mm (1/4") internal diameter rubber or plastic
hoses from the water supply must be fitted to the
two nozzles.
NOTE
These hoses must be held on the respective noz-
zles using hose clips to avoid that the tube(s) gets
loose or disconnected during operation.
The plastic model must be screwed into the suit-
able holes of the pump.
Cooling may be carried out either through an open
circuit with eventual discharge of the water, or us-
ing a closed circuit cooling system.
The water temperature must be between +10°C
and +30°C, with an inlet pressure between 3 and 5
bar. This allows a flow of about 200 l/h.

VENT VALVE INSTALLATION

The vent valve mod. 969-9833 allows to avoid un-
desired venting of the pump during a temporary
power failure (5 sec maximum), and enables an
automatic vent operation.
Unscrew the threaded plug (see figure below).
Screw the vent valve into the pump without insert-
ing the supplied adapter between the valve and the
pump.
Tighten the valve using a 10 mm. hexagonal span-
ner with a torque of 2.5 Nm.
! ! ! !
Do not overtighten the valve as this may damage
the thread on the pump.
30
TECHNICAL INFORMATION
CAUTION
87-900-897-01(C)
Tabla de contenido
loading

Este manual también es adecuado para:

969-9076969-9083969-9095969-9096

Tabla de contenido