20-22
Message/Symptom
Display: "- - -"
HAP 200 is not working
Display: "tip"
No air at HAP
No vacuum at unsoldering tool Vacuum hose not or
Status indication of VAC LEDs
incorrect
No display function (display off) No mains supply voltage
VAC LED red
10 Fault messages and fault elimination
Possible cause
Tool has not been detected
Tool defective
HAP 200 not connected to
channel 1
Soldering tip of microtool not
correctly inserted or defective
Air hose not or incorrectly
connected
incorrectly connected
Unsoldering nozzle clogged
Pressure-gauge level not
correctly set
Vacuum system clogged
Corrective measures
Check connection of tool to
device
Check connected tool
Connect HAP 200 to channel 1
Insert soldering tip again
Replacing defective soldering tip
Connect air hose to AIR nipple
Connect vacuum hose to Vac
nipple
Maintain unsoldering nozzle with
cleaning tool
Set pressure-gauge level in special
menu 2
Turn on mains power switch
Check mains supply voltage
Check device fuse
Clean suction nozzle
Check filter; replace if yellow
Clean unsoldering tool – replace
filter
Check vacuum hose
WR 3ME