Maintenance And Cleaning - Invacare Jasmine Manual Del Usuario

Ocultar thumbs Ver también para Jasmine:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

7.

Maintenance and cleaning

At normal daily operation, a service check-up should take place every year, according to the
maintenance chart.
When performing annual or regular maintenance, all parts designed to carry load must be as a
minimum tested with maximum load. All safety features must be checked according to EN ISO
10535: 2006 Annex B.
LOLER statement.
The UK Health and Safety Executive's Lifting Operations and Lifting Equipment Regulations 1998,
require any equipment that is used in the workplace to lift a load be subject to safety inspection
on a six monthly basis. Please refer to the HSE web site for guidance www.hse.gov.uk
It is the responsibility of the person responsible for the equipment to ensure that LOLER
regulations are adhered to.
The hoist must be stored at normal room temperature. If it is stored in a damp, cold or wet
environment then the motor and other mounting parts may be prone to corrosion.
The hoist
has been designed so that it requires a minimum amount of maintenance.
It is recommended that the batteries are charged frequently, if possible every night, as it will
extend the lifetime of the batteries and will ensure high performance.
It is recommended that the hoist
Please check that the spreader bar connections, control unit and motors are in good working
order and not damaged in any way. If there is any damage to the hoist, refrain from using it,
and contact your local Invacare
Wear and damage of loading parts
It is necessary to maintain and check all parts that are exposed to static or dynamic strain, such
as the sling, the spreader bar and the turning points, do not have fractures, are fragile, askew or
damaged.
Damaged parts must be replaced before further use.
Cleaning
To prevent cross-infection, the hoist must be cleaned after use.
The hoist must be wiped with a moist firmly wrung cloth with ordinary household
disinfectants.
Only use officially approved disinfection detergents.
Dry the hoist carefully after cleaning.
Never use acids, alkaline or solvents for cleaning the hoist.
Motors, control unit and mounting parts can be destroyed if the hoist is cleaned in any other way
than stated above.
is cleaned after use.
dealer before further use.
®