Placing and Replacing of the Interchangeable Unit

2.7.2
The base unit comes with an RFID reader, and all the interchangeable units are equipped with an RFID
transponder. This transponder can be used to store the following information:
Unit size (cannot be changed)
Unit ID (cannot be changed)
Reagent name (default: blank)
Concentration (default: 1.000000)
Concentration determined on: (Date)
To be used until: (Date)
Opened/Produced on: (Date)
Test according to ISO 8655: (Date)
Charge description: (default: no charge)
Last modification: (Date)
Each time an interchangeable unit is pushed onto the base unit, the data is automatically read out of the
transponder.
2.7.2.1 Placing an Interchangeable Unit
The interchangeable unit is to be placed on the device unit as is shown in Fig. 10 - Fig. 12 subsequently, it is to
be pushed downwards until the black button latches on the left side.
Fig. 10
Fig. 11
Fig. 12