Connection
14
When using the six‐wire technique, the effects of resist­
ance changes in the extension cable are compensated. If
you extend the cable with the four-wire technique, the
sensitivity deviation can be rectified by adjustment. Tem­
perature effects though are only compensated with the
operation using the six-wire technique.
The load cell connecting cable should be routed so that
any condensed water or dampness forming on the cable
can drip off. lt must not be led to the load cell. In addition,
it must be ensured that no dampness can penetrate the
open end of the cable.