4.1. MODE / PRINT KEY
4.1.1. PRIMARY FUNCTION BEFORE THE MEASUREMENT
The primary function of the MODE key is very important: it is used before the measurement to
defi ne the course of the measurement.
This key is inactive in the "Adjust. Step" and SET-UP positions.
Pressing the MODE key gives access to the list of possible measurement modes. Select the mode using
the
and/or
keys.
To validate the mode selected, press the MODE key again.
The various measurement modes are as follows:
MANUAL STOP:
This is the conventional quantitative insulation
measurement mode. The measurement is started
by pressing START / STOP and stopped by pressing
START / STOP again.
The user determines the duration, which is indicated
by the measurement duration chronometer.
MANUAL STOP + DD:
The measurement is started by pressing START/
STOP and stopped by pressing START/STOP again.
One minute after the end of the measurement, the
instrument calculates and display the DD term. The
time remaining during this minute is displayed.