DJO Global ARTROMOT-E2 Manual De Instrucciones página 35

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 77
Warning!
m
shock hazard —
strictly observe the following warnings.
Failure to do so endangers the lives of
the patient, the user and other persons
involved.
− Before use allow the ARTROMOT®-
E2/-E2 compact to reach room
temperature. if the device has been
transported at temperatures below
0 °C (32°F), leave it to dry at room
temperature for about 2 hours, until
any condensation has disappeared.
– The ARTROMOT®-E2/-E2 compact
must only be operated in dry rooms.
− when disconnecting the device from
the power line, remove the plug from
the wall outlet irst, before discon-
necting the cable from the device.
− when connecting the device to other
equipment or when creating a medi-
cal system, check that the sum of
leakage currents will not cause any
hazard. Please contact djO Global, if
you have questions in this matter.
– do not use multiple portable socket
outlets (MPsO) to connect the device
to the power line. The ARTROMOT®-
E2/-E2 compact must be connected
to a properly installed wall outlet
with a non-fused earthed wire.
before connecting the power cord,
it must be completely unrolled and
placed such that it will not get caught
in the moving parts of the device.
– before cleaning and service interven-
tions, disconnect the device from
the power line by removing the
power cord from the wall outlet.
– Liquids must not be allowed to en-
ter the CPM device or the program-
ming unit. if liquids have entered
into the device, the ARTROMOT®-
E2/-E2 compact must be checked by
a service technician, before it can be
reused.
60
Warning!
m
Equipment malfunction —
− Magnetic and electrical ields are
capable of interfering with the proper
performance of the device. For this
reason make sure that all external
devices operated in the vicinity of the
CPM device comply with the relevant
EMC requirements. X-ray equip-
ment, MRi devices, radio systems, cell
phones, etc. are possible sources of
interference as they may emit higher
levels of electromagnetic radiation.
keep the CPM device away from
these devices and verify its perform-
ance before use.
− Refer repair and maintenance to
authorized persons.
Persons are authorized after training
by a specialist trained and commis-
sioned by the manufacturer.
− Route all cables such that they will
not get caught by the moving parts
during operation and that they do
not present a stumbling hazard.
− Inspect the ARTROMOT®-E2/-E2
compact for damage and loose con-
nections at least once a year.
Caution!
m
Preventing chaing and pressure
sores —
– if your patient is adipose, very tall or
very short, be sure to prevent chaing
and pressure sores.
m
Caution!
Patient hazard, damage to the
CPM device —
– The CPM device ARTROMOT®-E2
must not be used to transport people.
Caution!
m
Equipment damage —
− Check that the voltage and frequency
ratings of your local power line are
those indicated on the nameplate.
− The seat withstands a maximum
continuous load of 150 kg / 330 lb.
− The arm support element with-
stands a maximum continuous load
of 9 kg / 20 lb.
− do not allow any objects (such as
blankets, cushions, or cables) to get
caught in the moving parts of the
CPM device.
− do not expose the ARTROMOT®-
E2/-E2 compact to direct sunlight,
because some of the components
may reach inadmissibly high tempe-
ratures.
Furthermore, the device must be set
up at a safe distance from radiators
to prevent excessive temperature
rises.
– The presence of children, pets and
rodents does not normally impair the
functioning of the device. however,
avoid contamination of the device by
children or animals, from dust and
lint, and keep them at a safe distance
from the device. The safety state-
ments set forth apply.
− be aware that the connectors can
only be inserted in the correct ori-
entation and secure all connections
with the locks.
61
Tabla de contenido
loading

Este manual también es adecuado para:

Artromot-e2 compact

Tabla de contenido