Table of contents
Installation
on page 7
Startup
on page 8
Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions (D x W x H)
18.5 x 33.0 x 40.0 cm (7.3 x 12.5 x 15.7 in.)
Weight
9.3 kg (20.5 lb)
Number of channels
8
Size channels
ISO-MTD 4, 4.6, 6, 9.8, 14, 21.2, 38, 68 μm
ACFTD ~1, 2, 5, 10, 25, 50, 100 μm
Flow rate
15 to 50 ml/min (automatic/manual)
Light source
Class 3B laser, 810-852 nm, 50 mW maximum
Calibration
ISO MTD (based on ISO 11171)
Full ISO 11171 Optional
Counting efficiency
JIS B9925:1997
Concentration limit
20,000 particles per ml @ 5% coincidence loss
(per ISO 11171)
30,000 particles/ml @ 10% coincidence
Sample volume
3 runs (averaged) of 5, 10 or 20 ml/run
(programmable)
User interface and navigation
on page 9
Operation
on page 12
Specification
Fluid temperature range
0 to 90 °C at 25 °C ambient (32 to 194 °F at
77 °F)
Measured fluid temperature
0 to 100 °C, ±0.5 °C (32 to 212 °F, ±0.9 °F)
Viscosity range
10 to 424 cSt (59 to 2000 SUS)
For viscosity measurement, viscosity range is
10 to 424 cSt ±20% at value.
Wetted materials
Aluminum, stainless steel, sapphire, PTFE and
Aflas
Cleanliness classification
ISO 4406-1991, ISO 4406-1999, NAS 1638, MIL-
STD-1246C, SAE AS 4059, NAVAIR 01-1A-1
Data storage
500 sample records
Input/output serial
RS232
communication
Altitude
2000 m (6561.7 feet)
Certifications
CE
Bottle operation
Purge volume
15 to 30 ml (automatic/manual)
Programmable
Cartridge
CO
, replaceable, rechargeable
2
Operating time
60 samples (120 ml sample bottle)
Programmable, hold time and sample number
Shop air
620 to 758 kPa (90 to 100 psi; 6.2 to 7.6 bar)
clean, dry
Fluid pressure
40 to 5000 psi (2.75 to 345 bar)
Power
DC input
24 VDC, 2 A maximum
AC adapter
Universal 100-240 VAC, 50-60 Hz, 60 W
Rechargeable battery
Nickel-metal hydride
®
English 3