5 Maintenance
To ensure that your weighing module remains reliable, accurate, and functional for many years to come, the
individual components must be cleaned and maintained periodically as appropriate for the intensity of use and
the risk of contamination with debris.

5.1 Cleaning the weigh module

Clean the weighing pan and the load cell housing regularly with a damp cloth. The electronic unit and terminal
can also be cleaned in this fashion as needed. For tougher dirt, a mild household cleaner may be used. Make
sure that no liquid penetrates inside the components!
Please observe the following notes:
Risk of electric shock
1 Disconnect the weigh module from the power supply prior to cleaning and maintenance.
2 Only use METTLER TOLEDO power cable, if these need to be replaced.
3 Ensure that no liquid comes into contact with the weigh module, terminal or AC adapter.
4 Do not open the weigh module, terminal or AC adapter.
Damage to weigh module
Under no circumstances use cleaning agents containing solvents or abrasive agents, as this can
damage or scratch certain surfaces (particularly the terminal screen).
You can also clean the load cell using a low-pressure hand sprayer.
Before using a sprayer:
1 Remove the weighing pan and the wind ring and seal off the
weighing pan retainer with the white plastic cover (1).
2 Make sure that the cover locks on properly!
3 Screw the connector cable to the electronic unit on.
4 Ensure that the jack (2) is sealed.
ð Under these conditions, the load cell meets IP45 requirements.
5 Dry the load cell it off with a soft cloth.
Note
The sealing function of the plastic cover may deteriorate over time with
frequent use. Check the condition of the cover before you use it. If
necessary, a new cover can be ordered as a replacement part see
Spare Parts in the separate Installation Manual.

5.2 Cleaning the wind shield (WXTS3DU)

1 Open the left and the right wind shield door and remove the base plate protection from the wind shield.
2 Remove the top glass.
3 Use a soft cloth to clean the parts of the wind shield housing.
Note
The elements of the wind shield can also be removed from the plate and be cleaned in a dish washer.
10
Maintenance​ ​
WARNING
These contain no user-serviceable parts.