6.1
REPLACING THE BATTERIES
The remaining battery charge level is indicated by the
When the battery is flat (the
beeps), then automatically switches to standby. When the battery is low, display of the "BAtt" message
indicates that the measurement requested requires too much energy and cannot be made.
Note: when rechargeable batteries are used, this fact must be entered in the configuration of the
instrument ("SET-UP" mode) to prevent incorrect operation of the instrument (risk of erroneous
measurements or malfunction of the instrument).
Check that none of the input terminals is connected and that the switch is set to OFF before
opening the instrument.
When the batteries are removed, there is enough reserve energy to preserve the date and time for one
minute. If this time is exceeded, the instrument, when next started up, prompts the user to check the
date and time by displaying the flashing message "tIME" for 2 seconds before displaying new
measurements.
6.2
STORAGE OF THE INSTRUMENT
If the instrument must be stored more than 2 months, remove the batteries. Remember to reset the
instrument to the correct time the next time you use it.
6.3
CLEANING
Clean the housing of the instrument regularly. This can be done with a damp cloth or soapy water. Do
not use alcohol, solvents, or hydrocarbons.
6.4
METROLOGICAL VERIFICATION
Like all measuring and testing devices, the instrument requires periodic verification.
We recommend checking the instrument at least once a year. For verifications and calibrations,
contact our COFRAC-accredited metrology laboratories or the MANUMESURE agencies.
Information and coordinates on request: Tel.: 02 31 64 51 43
6.5
CUSTOMER SERVICE
Notre garantie s'exerce, sauf stipulation expresse, pendant 12 mois après la date de mise à disposition
du matériel (extrait de nos Conditions générales de Vente, communiquées sur demnde).
6.6
SERVICE APRÈS-VENTE
For maintenance work, use only specified spare parts.
The manufacturer cannot be held liable for any accident that occurs following a repair done
other than by its Customer Service Department or an approved repairer.
Repair under and out of warranty:
Send the instrument to one of the MANUMESURE regional agencies, approved by Chauvin Arnoux
Information and coordinates on request:
Repair outside mainland France:
For any repair (under or out of warranty), send the instrument back to the dealer.
6. MAINTENANCE
symbol flashes), the instrument emits an audible stop signal (5
Tel.: 02 31 64 51 43 - Fax: 02 31 64 51 09
75
symbol.
-
Fax: 02 31 64 51 09