The TP650/TP700 is suitable for use in the electromagnetic environment specified below.
The customer or the user of the TP650/TP700 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1
At 80 MHz and 800 Mhz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radiobroadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the TP650/TP700 is used exceeds the applicable RF compliance level above, the TP650/TP700 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the TP650/TP700.
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m
Guidance and manufacturer's declaration – electromagnetic emissions
The TP650/TP700 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TP650/TP700 should assure that it is used in such an environment
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
www.stryker.com
EMC Information
TP650/TP700 EMC Information
Compliance
level
Portable and mobile RF communications equipment should be used
no closer to any part of the
3 Vrms
the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
d = 1,2
3 V/m
d = 1,2
d = 2,3
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
Should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
P
Compliance
Electromagnetic Environment
Group 1
The TP650/TP700 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The TP650/TP700 is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage power
supply network that supplies buildings used for domestic purposes.
Class A
Complies
101515 REV C
Electromagnetic environment - guidance
TP650/TP700
, including cables, than
Recommended Separation Distance
P
80 MHz to 800 MHz
P
800 MHz to 2,5 GHz
a
b
1-23