Gima 32597 Manual Del Usuario página 79

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

79
• The doctor interface is as Fig. 6-22.
You can set the name, telephone and address of doctors in this interface.
You can input and record 10 doctor names at most, and it relates with
the diagnosis doctor item of input case dialog box.
Fig. 6-22
• In the "sampling setting" (Fig. 6-23), you can set the power frequency filter
parameters, EMG Filter, Base Line Filter, Device Type and so on.
Fig. 6-23
Set the power frequency filter parameters. The "Frequency" item must be set
to the power frequency of your country or region. And it is 50 Hz in China.
If "Base Line Filter" and "EMG Filter" are selected, the ECG wave will be better,
but at the same time they will affect the ECG signal to some extent.
So we propose to choose them cautiously.
Set the device version.The "Device Type" item must be set to the version of your
device. And the default version is "Without Blueth".
The choices you set will be applied to actual sampling automatically and
regarded as the default, but it does not mean that we can't change the settings
during sampling. You can also change them manually before or after sampling.
ENGLISH
Tabla de contenido
loading

Este manual también es adecuado para:

Cms-vesd

Tabla de contenido