48
Elektromagnetische
Verträglichkeit EMV
Recommended separation distances between portable and mobile RF telecommunication devices and
The ARTIP™ BASE holding system 28272 TLS and 28272 TSS is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or user of the ARTIP™ BASE holding system 28272 TLS and 28272 TSScan help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ARTIP™ BASE holding system
28272 TLS and 28272 TSS as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
nOtE: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
nOtE: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of
structures, objects and people.
Compatibility with surgical HF instruments
This device was tested for compatibility with HF surgical devices.
Electromagnetic
Compatibility EMC
Table 5
the ARTIP™ BASE holding system 28272 TLS and 28272 TSS
Separation distance d [m]
as a function of the transmitter frequency
150 kHz to 80 MHz
d
0.12
0.38
1.2
3.8
12
Compatibilidad
electromagnética CEM
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d
0.12
0.38
1.2
3.8
12
e
0.23
0.73
2.3
7.3