The SUPER TOBI suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and always check integrity of plastic parts and feeding
cable, they might have been damaged during previous use. Connect cable to electrical network and turn switch on. Close the aspirator outlet with your
finger and with suction regulator in maximum vacuum position check that the vacuum indicators reaches -80kPa (-0.80 bar) minimum.
Rotate the knob from right to left and check the aspiration regulating control. The vacuum indicator should go down -40kPa (-0.40 bar). Verify that loud
noises are not present, these can indicate wrong functioning. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situated in the plug
protects the instrument. For fuses replacing, always the type and the range.
Type of fault
1. The suction unit doesn't work
2. No aspiration
3. No aspiration
4. The Vacuum power on the patient
side is either very low or absent
5. The float doesn't close
6. The float doesn't close
7. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7
If the overfill security system it's activated, don't proceede with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid penetration inside the
device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the device to GIMA
technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT GIMA
TECHNICAL SERVICE.
GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE
TAMPERED.
• Connect the short silicon tube, with antibacterial filter, to the suction connector. The other tube, with one end connected to the filter must
be connected with the other end to jar's lid connector where has been fixed the red float.
When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on
the jar) to avoid liquid penetration inside the device. The device must be used on a plan of horizontal operation.
WARNING: Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the "VACUUM".
A wrong connection causes immediate destruction in case of contact with sucked liquids.
• Connect the long silicon tube to the other jar's lid connector
• Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it.
• Connect the power cord to the device then connect the plug to the electrical mains supply.
• Push switch on position I to start suction
• Once finished push switch on 0 position and unplug.
• Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the functionally
vacuum) then rescrew the lid on the jar correctly.
• To extract the accessories and start with cleaning.
PERIODICAL MAINTENANCE CHECKS
Cause
Cable is damaged
External power source failure
Jar Cap badly screwed down
Lid seal not in its seat
• Vacuum regulator set to minimum
• Protection filter blocked or damaged
• Connection tubes blocked, kinked or
disconnected
• Shut-off valve blocked or damaged
• Pump motor damaged
If the cap has been washed, ensure that
the float is not partially detached
The float it's covered by dirty material
Foam inside the jar
None of the remedies has achieved the
desired results
INSTRUCTION FOR USE
Remedy
Replace the cable
Check the external power source
Unscrewed the cap, then rescrew it correctly
Unscrew the cap and insert the seal properly in its seat
• Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
• Replace the filter
• Replace or reconnect the tubes, check the jar
connections
• Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in
the upright position
• Refer to authorised service personnel
Insert the float into it's place
Unscrewed the cap, leave the and put in on autoclave
Fill the jar to 1/3 full of ordinary water
Contact the seller or GIMA After-sales Assistance