To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the jar from the device
Disconnect all tubes from the jar and the protection filter
Empty and dispose of the content and of the suction catheter according to the laws in force in your country;
Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly. Then soak in warm water (temperature
shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to remove incrustations.
Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).The jar and the cover can be autoclaved by placing the parts into the
autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit
the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
Assemble the jar as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C (1 bar relative pressure – 15 min)
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C. (1 bar relative pressure – 15 min)
The device is ready for a new employment now.
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit. When the filter is wet, it's not possible to use
the unit therefore the filter should be changed immediately. In case of possible contamination or discolouration, change the filter immediately.
Don't use the suction unit without the protection filter fitted. If the suction unit is used in an emergency or in a patient where the risk of contamination is not
know the filter must be changed after each use. Available under request with different versions with complete jar 2000cc.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-infections.
Don't use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standards
on material biocompatibility.

CLEANING OF ACCESSORIES

ACCESSORIES SUPPLIES

DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6mm x 10mm (TRASPARENT SILICON)
ASPIRATION PROBE CH 20
ANTIBACTERIAL FILTER